RESPUESTA INFLAMATORIA AL BY-PASS CARDIOPULMONAR: UNA REVISIÓN TERAPÉUTICA. |
| Rubens FD, Mesana T. |
|
Division of Cardiac Surgery, University of Ottawa Heart Inatitute, Ottawa, Ontario, Canada. |
| Trabajo original publicado en Perfusion 19: S5-S12, 2004. |
| Traducido con permisión de los editores. Traducido por Juan Carlos Vázquez. |
|
|
The demographic of cardiac surgery patients continues to
evolve to include older, sicker candidates, all the while
maintaining an expectation of excellent outcomes. These
latter results can only be achieved by the parallel
advancement and re-examination of the technology of
cardiopulmonary bypass (CPB); the key tool used daily by
surgical teams worldwide. In this review, we will provide
an overview of integrated therapeutic strategies that can
be utilized to minimize the complex and myriad changes
related to inflammation after CPB with the understanding
that this may abrogate the detrimental end-organ and
systemic effects of blood activation. Therapeutic strategies
specifically related to the technology can be classified
into those targeting biomaterial dependent or
independent processes. The former can be addressed
by the utilization of currently available biocompatible
surfaces such as with heparin-coated circuits, phosphorylcholine-
coated circuits (‘biomembrane mimicry’) and
circuits composed of copolymers containing surfacemodifying
additives. The most important strategies related
to biomaterial independent activation include the
modification of techniques related to cardiotomy blood
management and blood filtration. Finally, all of these
strategies must be integrated and tailored with complementary
pharmacologic agents such as aprotinin and
steroids to optimize anti-inflammatory synergism. Only if
we are armed with a comprehensive knowledge of the
molecular and cellular basis for these strategies will we
be able to continue to evolve our treatment in parallel
with our patients to achieve these goals.
Rev Latinoamer Tecnol Extracorp 12,2,2005
Las características demográficas de la población sometida a cirugía cardíaca, continúa evolucionando incluyendo pacientes más añosos y con mayor enfermedad, manteniendo una expectativa de obtener excelentes resultados.
Estos resultados alejados pueden ser solamente obtenidos con una re-examinación simultánea de las técnicas de circulación extracorpórea (CEC).
En esta revisión, proveeremos una síntesis de las estrategias terapéuticas integradas que pueden ser empleadas para minimizar el complejo inflamatorio después de CEC, con el convencimiento que dichas estrategias pueden ayudar a evitar la disfunción orgánica y los efectos sobre la activación de los componentes sanguíneos.
Las estrategias terapéuticas referentes a esta tecnología pueden ser clasificadas en aquellas dependientes de los biomateriales empleados y aquellas que dependen de procesos independientes.
Las primeras pueden ser expresadas por la utilización de superficies biocompatibles actualmente disponibles, como circuitos recubiertos con heparina, fosforilcolina y circuitos conteniendo copolímeros que modifican la superficie de los circuitos. Las técnicas más importantes relativas a los factores independientes incluyen la modificación de la aspiración de sangre
del campo operatiorio y la filtración de sangre. Finalmente, ambas técnicas pueden ser conjugadas con el apoyo complementario de agentes farmacológicos como la aprotinina y los esteroides para optimizar el sinergismo anti-inflamatorio.
Solamente si nos encontramos consustanciados con la comprensión de los fenómenos celulares y moleculares que comprenden estas estrategias, estaremos en condiciones de desarrollar nuestra acción en paralelo a la complejidad de nuestros pacientes, y podemos alcanzar nuestros objetivos.
INTRODUCCION
Las complejas preguntas relativas a la respuesta inflamatoria sistémica post CEC, han sido suplantadas por los tópicos científicos alternativos al uso de tecnologías en CEC. El más notable entre ellos es el uso de cirugía coronaria sin CEC (OPCAB) que alternativamente aumenta o disminuye sus adeptos.
Para un observador casual esta técnica parece atractiva, sin embargo las limitaciones y contraindicaciones del OPCAB, reafirman la necesidad de continuar investigando para el mejoramiento de la técnica de perfusión. Ahora y en el futuro, nos encontraremos con una población más enferma, con inestabilidad hemodinámica y enfermedad de arterias pequeñas que descartan las técnicas de OPCAB así como expanden la población que necesita perfusión como los reemplazos valvulares.
De hecho, es en este grupo de alto riesgo, que se observarán los beneficios de la investigación sobre biocompatibilidad y los esfuerzos que se realicen para mejorar las técnicas de CEC.
Sin duda, la respuesta inflamatoria sistémica a la cirugía cardíaca es influenciada por factores generales del ''trauma controlado'', con múltiples incisiones y el impacto sistémico de los agentes anestésicos. Esta revisión se concentra en aquellos aspectos modificables del proceso como el tipo de equipamientos y las tácticas operativas.
En primer término, nosotros y otros autores hemos provisto evidencias que las interacciones de la sangre con los biomateriales dependiendo de su constitución, pueden tener un rol significativo en la respuesta humoral sistémica a la CEC (biomateriales-dependiente) y esta elección puede jugar un rol importante en la evolución de los pacientes, especialmente en aquellos casos de alto riesgo [1]. Una clasificación del espectro de tecnología disponible actualmente, puede ayudarnos a reconocer el significado por el cual pueden mejorar esta respuesta.
La otra estrategia (biomaterial-independiente) se refiere a las alternativas de manejo como el tratamiento de la sangre aspirada por cardiotomía, el uso de agentes farmacológicos que disminuyen la respuesta y distintos materiales (Ej.: filtros) que previamente no habían sido reconocidos como un importante enfoque para limitar la reacción inflamatoria en la población sometida a cirugía cardíaca adulta.
FACTORES DEPENDIENTES DE LOS BIOMATERIALES DURANTE LA CEC
Existe una compleja mirada para comprender la interacción de los biomateriales con la sangre, y el lector debe referirse a numerosas fuentes y revisiones para tener perspectiva sobre el tema [2]. Sin embargo, la cuestión central para calificar los fenómenos que mejoran la biocompatibilidad, puede ser categorizada en tres procesos dinámicos:
El primero refiere a una relativa resistencia de las proteínas a activarse al contacto con los materiales. El segundo se refiere a la activación del complemento. A pesar que este proceso puede ser secundario a la adsorción de las proteínas, particularmente las inmunoglobulinas, esto se complica por la interacción de la activación de la kalicreína y la liberación de citoquinas [3,4]. Finalmente, un predictor sensitivo de biocompatibilidad se refiere a la dinámica de la adsorción [5] del fibrinógeno, no sólo como resultado de su densidad absoluta [6,7], sino por la naturaleza de los cambios que pueden ocurrir después de su adsorción. Hay una gran interrelación entre este evento y el grado de acumulación de plaquetas en la superficie [8].
A pesar del rol predominante de las plaquetas y el fibrinógeno en el proceso de trombosis [9], se deberán reconocer estos pasos y relacionarlos estrictamente con la activación inflamatoria siendo prácticamente imposible diferenciar los procesos de trombosis e inflamación desde un punto de vista estrictamente académico.
ALTERNATIVAS EN BIOMATERIALES PARA CIRCUITOS DE CEC
Los factores económicos han limitado la evolución de nuevos materiales para CEC, de manera que actualmente, permanecen solo tres tipos primarios de superficies biocompatibles disponibles comercialmente para uso clínico.
El primero de ellos se distingue por la estrategia '' biomembrana mímica'' capitalizando el conocimiento que la tromboresistencia de las membranas biológicas nativas no activadas (Ej.: eritrocitos) es la llave para mantener la circulación fluída. La fosforilcolina, el componente fosfolípido que se encuentra en ambos bordes de la superficie es eléctricamente neutro, en cambio su contraparte la fosforilserina ubicada en la superficie intermedia mantiene su polaridad trombogénica [10].
La fosforilcolina ha sido incorporada dentro de un copolímero( methacrylophosphorilcholine-laruyl methacrylate,MPC:LM) el cual tiene propiedades hidrofílicas y se refiere como un hidrogel [11,12].
El hidrogel es relativamente sencillo para cubrir cualquier superficie con polímeros post-fabricación de manera que se mantiene estable, segura y con mínimo desprendimiento.
Esta superficie ha demostrado ser durable manteniendo la integridad de la misma por largos períodos de tiempo. Stents coronarios, recubiertos con un proceso similar, han sido explantados luego de más de seis meses manteniendo intacta la malla MPC:LM tanto en animales [13] como en humanos [14].
In vitro, la MPC:LM cobertura ha demostrado resistencia a la adhesión proteica en particular al fibrinogel probablemente debido a las propiedades intrínsecas hidrofílicas de esta superficie (figura 1).
|
| Figura 1. Cubierta de policarbonato y PVC con MPC:LM han causado una significante redución en la adsorción del fibrinógeno. |
|
| Figura 2. Local ambiental de las superficies hidrofílico-hidrofobicas en contacto con los componentes de la sangre, incluso celulas y proteinas. |
|
| Figura 3. Marcadores de la generación de trombina en los tests y en los controles. |
1. Rubens FD, Labow RS, Lavallee GR et al. Hematologic evaluation of cardiopulmonary bypass circuits prepared with a novel block copolymer. Ann Thorac Surg 1999; 67: 689 ¡/98.
2. Rubens FD. Cardiopulmonary bypass: technique and pathophysiology. In Selke FW ed. Sabiston & Spencer surgery of the chest, seventh edition, New York, NY: Elsevier, 2004.
3. Heller W, Fuhrer G, Gallimore MJ, Michel J, Hoffmeister HE. Changes in the kallikrein-kinin-system after different dose regimen of aprotinin during cardiopulmonary bypass operation. Adv Exp Med Biol 1989; 247B: 43¡/48.
4. Ichinose A, Kisiel W, Fujikawa K. Proteolytic activation of tissue plasminogen activator by plasma and tissue enzymes. FEBS Lett 1984; 175: 412 ¡/8.
5. Tengvall P, Askendal A, Lundstrom I. Temporal studies on the deposition of complement on human colostrum IgA and serum IgG immobilized on methylated silicon. J Biomed Mater Res 1997; 35: 81¡/92.
6. van Deventer SJ, Hack CE, Wolbink CE et al . Endotoxin- induced neutrophil activation¡/the role of complement revisited. Prog Clin Biol Res 1991; 367: 101 ¡/9.
7. Steinberg JB, Kapelanski DP, Olson JD, Weiler JM. Cytokine and complement levels in patients undergoing cardiopulmonary bypass. J Thorac Cardiovasc Surg 1993; 106: 1008 ¡/16.
8. Rubens FD, Brash JL, Weitz JI, Kinlough-Rathbone RL. Interactions of thermally-denatured .brinogen on polyethylene with plasma proteins and platelets. J Biomed Mater Res 1992; 26: 1651 ¡/63.
9. Sheppeck RA, Bentz M, Dickson C et al . Examination of the roles of glycoprotein Ib and glycoprotein IIb/IIIa in platelet deposition on an arti.cial surface using clinical antiplatelet agents and monoclonal antibody blockade. Blood 1991; 78: 673 ¡/80.
10. Campbell EJ, O’Byrne V, Stratford PW et al . Biocompatible surfaces using methacryloylphosphorylcholine laurylmethacrylate copolymer. ASAIO J 1994; 40: M853 ¡/57.
11. Sugiyama K, Matsumoto T, Yamazaki Y. Evaluation of biocompatibility of the surface of polyethylene .lms modi.ed with various water soluble polymers using Ar plasma ¡/ post polymerization technique. Macromol Mater Eng 2000; 282: 5¡/12.
12. Murphy EF, Lu JR, Brewer J, Russell J, Penfold J. The reduced adsorption of proteins at the phosphoryl Inflammation and cardiopulmonary bypass FD Rubens and T Mesana S10 choline incorporated polymer¡/water interface. Langmuir 1999; 15: 1313¡/22.
13. Lewis AL, Tolhurst LA, Stratford PW. Analysis of a phosphorylcholine-based polymer coating on a coronary stent pre- and post-implantation. Biomaterials 2002; 23: 1697 ¡/706.
14. Lewis AL, Furze JD, Small S et al . Long-term stability of a coronary stent coating post-implantation. J Biomed Mater Res 2002; 63: 699 ¡/705.
15. De Somer F, Van Belleghem Y, Caes F et al . Tissue factor as the main activator of the coagulation system during cardiopulmonary bypass. J Thorac Cardiovasc Surg 2002; 123: 951 ¡/8.
16. De Somer F, Francois K, van Oeveren W et al . Phosphorylcholine coating of extracorporeal circuits provides natural protection against blood activation by the material surface. Eur J Cardiothorac Surg 2000; 18: 602¡/6.
17. TrilliumTM Af.nity NT Oxygenator. Product monograph, 2000.
18. Elgue G, Blomback M, Olsson P, Riesenfeld J. On the mechanism of coagulation inhibition on surfaces with end point immobilized heparin. Thromb Haemost 1993; 70: 289 ¡/93.
19. Ovrum E, Holen EA, Tangen G et al . Completely heparinized cardiopulmonary bypass and reduced systemic heparin: clinical and hemostatic effects. Ann Thorac Surg 1995; 60: 365 ¡/71.
20. Aldea GS, Doursounian M, O’Gara P et al. Heparinbonded circuits with a reduced anticoagulation protocol in primary CABG: a prospective, randomized study. Ann Thorac Surg 1996; 62: 410 ¡/18.
21. Mahoney CB. Heparin-bonded circuits: clinical outcomes and costs. Perfusion 1988; 13: 192¡/204.
22. te Velthuis H, Baufreton C, Jansen PGM et al . Heparin coating of extracorporeal circuits inhibits contact activation during cardiac operations. J Thorac Cardiovasc Surg 1997; 114: 117 ¡/22.
23. Baufreton C, Jansen PG, Le Besnerais P et al . Heparin coating with aprotinin reduces blood activation during coronary artery operations. Ann Thorac Surg 1997; 63: 50¡/56.
24. Gorman RC, Ziats NP, Rao AK et al. Surface-bound heparin fails to reduce thrombin formation during clinical cardiopulmonary bypass. J Thorac Cardiovasc Surg 1996; 111: 1¡/12.
25. Ovrum E, AmHolen E, Tangen G, Ringdal M-AL. Heparinized cardiopulmonary bypass and full heparin dose marginally improve clinical performance. Ann Thorac Surg 1996; 62: 1128 ¡/33.
26. Ranucci M, Mazzucco A, Pessotto R et al. Heparincoated circuits for high-risk patients: a multicenter, prospective, randomized trial. Ann Thorac Surg 1999; 67: 994¡/1000.
27. Khuri S, Valeri CR, Loscalzo J et al . Heparin causes platelet dysfunction and induces .brinolysis before cardiopulmonary bypass. Ann Thorac Surg 1995; 60: 1008 ¡/14.
28. Muriithi EW, Belcher PR, Day SP, Menys VC, Wheatley DJ. Heparin-induced platelet dysfunction and cardiopulmonary bypass. Ann Thorac Surg 2000; 69: 1827 ¡/ 32.
29. Chenoweth DE, Cooper SW, Hugli TE, Stewart RW, Blackstone EH, Kirklin JW. Complement activation during cardiopulmonary bypass: evidence for generation of C3a and C5a anaphylatoxins. N Engl J Med 1981; 304: 497¡/503.
30. Olson ST, Sheffer R, Francis AM. High molecular weight kininogen potentiates the heparin-accelerated inhibition of plasma kallikrein by antithrombin: role for antithrombin in the regulation of kallikrein. Biochemistry 1993; 32: 12136 ¡/47.
31. Deppisch R, Gohl H, Smeby L. Microdomain structure of polymeric surfaces ¡/ potential for improving blood treatment procedures. Nephrol Dial Transplant 1998; 13: 1354 ¡/9.
32. Deppisch R, Storr M, Buck R, Go˜hl H. Blood material interactions at the surfaces of membranes in medical applications. Separation Purif Technol 1998; 14: 241¡/ 54
33. Okano T, Aoyagi T, Kataoka K et al . Hydrophilic¡/ hydrophobic microdomain surfaces having an ability to suppress platelet aggregation and their in vitro antithrombogenicity. J Biomed Mater Res 1986; 20: 919¡/ 27.
34. Defraigne JO, Pincemail J, Dekoster G et al . SMA circuits reduce platelet consumption and platelet factor release during cardiac surgery. Ann Thorac Surg 2000; 70: 2075 ¡/81.
35. Suhara H, Sawa Y, Nishimura M et al . Ef.cacy of a new coating material, PMEA, for cardiopulmonary bypass circuits in a porcine model. Ann Thorac Surg 2001; 71: 1603¡/8.
36. Fearon DT. Identi.cation of the membrane glycoprotein that is the C3b receptor of the human erythrocyte, polymorphonuclear leukocyte, B lymphocyte, and monocyte. J Exp Med 1980; 152: 20¡/30.
37. Andersen LW, Landow L, Baek L, Jansen E, Baker S. Association between gastric intramucosal pH and splanchnic endotoxin, antibody to endotoxin, and tumor necrosis factor-alpha concentrations in patients undergoing cardiopulmonary bypass. Crit Care Med 1993; 21: 210 ¡/17.
38. Oudemans-van Straaten HM, Jansen PG, Hoek FJ et al . Intestinal permeability, circulating endotoxin, and postoperative systemic responses in cardiac surgery patients. J Cardiothorac Vasc Anesth 1996; 10: 187¡/ 94.
39. Schonberger JP, Everts PA, Hoffman JJ. Systemic blood activation with open and closed venous reservoirs. Ann Thorac Surg 1995; 59: 1549¡/55.
40. de Haan J, Boonstra PW, Monnink SHJ, Ebels T, van Oeveren W. Retransfusion of suctioned blood during cardiopulmonary bypass impairs hemostasis. Ann Thorac Surg 1995; 59: 901 ¡/7.
41. Chung JH, Gikakis N, Rao K et al . Pericardial blood activates the extrinsic coagulation pathway during clinical cardiopulmonary bypass. Circulation 1996; 93: 2014 ¡/8.
42. Tabuchi N, de Haan J, Boonstra PW, van Oeveren W. Activation of .brinolysis in the pericardial cavity during cardiopulmonary bypass. J Thorac Cardiovasc Surg 1993; 106: 828 ¡/33.
43. Rucinski B, Niewiarowski S, James P, Walz DA, Budzynski AZ. Antiheparin proteins secreted by human platelets. Puri.cation, characterization, and radioimmunoassay. Blood 1979; 53: 47¡/62.
44. Weitz JI, Hudoba M, Massel D, Maraganore J, Hirsh J. Clot-bound thrombin is protected from inhibition by heparin-antithrombin III but is susceptible to inactiva- Inflammation and cardiopulmonary bypass FD Rubens and T Mesana S11 tion by antithrombin III-independent inhibitors. J Clin Invest 1990; 86: 385 ¡/91.
45. Parratt R, Hunt BJ. Direct activation of factor X by monocytes occurs during cardiopulmonary bypass. Br J Haemat 1998; 101: 40¡/46.
46. Brooker RF, Brown WR, Moody DM et al . Cardiotomy suction: a major source of brain lipid emboli during cardiopulmonary bypass. Ann Thorac Surg 1998; 65: 1651 ¡/55.
47. Andersen LW, Baek L, Thomsen BS, Rasmussen JP. Effect of methylprednisolone on endotoxemia and complement activation during cardiac surgery. J Cardiothorac Anesth 1989; 3: 544 ¡/9.
48. Wan S, LeClerc JL, Huynh CH et al . Does steroid pretreatment increase endotoxin release during clinical cardiopulmonary bypass? J Thorac Cardiovasc Surg 1999; 117: 1004 ¡/8.
49. Teoh KH, Bradley CA, Gauldie J, Burrows H. Steroid inhibition of cytokine-mediated vasodilation after warm heart surgery. Circulation 1995; 92: II347¡/53.
50. Hill GE, Snider S, Galbraith TA, Forst S, Robbins RA. Glucocorticoid reduction of bronchial epithelial in- ammation during cardiopulmonary bypass. Am J Respir Crit Care Med 1995; 152: 1791 ¡/5.
51. Kawamura T, Inada K, Nara N, Wakusawa R, Endo S. Inuence of methylprednisolone on cytokine balance during cardiac surgery [see comments] [published erratum appears in Crit Care Med 1999; 27: 1404]. Crit Care Med 1999; 27: 545 ¡/8.
52. Hill GE, Alonso A, Thiele GM, Robbins RA. Glucocorticoids blunt neutrophil CD11b surface glycoprotein upregulation during cardiopulmonary bypass in humans. Anesth Analg 1994; 79: 23¡/27.
53. Jansen NJ, van Oeveren W, van Vliet M et al . The role of different types of corticosteroids on the inammatory mediators in cardiopulmonary bypass. Eur J Cardiothorac Surg 1991; 5: 211 ¡/7.
54. Kawamura T, Inada K, Okada H, Okada K, Wakusawa R. Methylprednisolone inhibits increase of interleukin 8 and 6 during open heart surgery. Can J Anaesth 1995; 42: 399¡/403.
55. Jansen NJ, van Oeveren W, van den Broek L et al . Inhibition by dexamethasone of the reperfusion phenomena in cardiopulmonary bypass [see comments]. J Thorac Cardiovasc Surg 1991; 102: 515 ¡/25.
56. Yilmaz M, Ener S, Akalin H et al. Effect of low-dose methyl prednisolone on serum cytokine levels following extracorporeal circulation. Perfusion 1999; 14: 201¡/6.
57. Mayumi H, Matsui K, Matsuzaki K et al . Cellular kinetics of posttransfusion graft-versus-host disease after heart operations [letter]. J Thorac Cardiovasc Surg 1994; 108: 179 ¡/82.
58. Mayumi H, Nakashima A, NishimiM et al . Risk factors for posttransfusion graft versus host disease, mediastinitis, and late cardiac tamponade in heart surgery. Survey of 119 Japanese institutions [In Process Citation]. Jpn J Thorac Cardiovasc Surg 2000; 48: 47¡/55.
59. Mayumi H, Zhang QW, Nakashima A et al . Synergistic immunosuppression caused by high-dose methylprednisolone and cardiopulmonary bypass. Ann Thorac Surg 1997; 63: 129 ¡/37.
60. Rich JB. The ef.cacy and safety of aprotinin use in cardiac surgery. Ann Thorac Surg 1998; 66: S6¡/S11.
61. Wachtfogel YT, Kucich U, Hack CE et al . Aprotinin inhibits the contact, neutrophil, and platelet activation systems during simulated extracorporeal perfusion. J Thorac Cardiovasc Surg 1993; 106: 1¡/9.
62. Kim KU, Kwon OJ, Jue DM. Pro-tumour necrosis factor cleavage enzyme in macrophage membrane/particulate. Immunology 1993; 80: 134 ¡/9.
63. Diego RP, Mihalakakos PJ, Hexum TD, Hill GE. Methylprednisolone and full-dose aprotinin reduce reperfusion injury after cardiopulmonary bypass. J Cardiothorac Vasc Anesth 1997; 11: 29¡/31.
64. Tassani P, Richter JA, Barankay A et al . Does high-dose methylprednisolone in aprotinin-treated patients attenuate the systemic inammatory response during coronary artery bypass grafting procedures? J Cardiothorac Vasc Anesth 1999; 13: 165 ¡/72.